2.1 Operating principles

Metrohm 838 Advanced Sample Processor, Operation

3

2 Operation

As well as the basic functions that can be carried out manually, this sec-

tion also described the configuration of the Advanced Sample Proces-

sor. Detailed descriptions of the way that run sequences are created and

how methods are managed are given.

2.1 Operating principles

2.1.1 Display

The keypad (SC Controller) display consists of two lines each with 24

characters.
The first line is used as the title line and shows the current method to-

gether with the number of processed samples. In the editing mode the

menu titles are shown here.
The second line is used as the status line and shows specific activities

that depend on the operating condition. It is used as the input line in the

editing mode.

Normal state

Sample

counter

Method name

Pump status

******** counter 1/12
PUMP--++STIR+---- ready

Instrument status

Stirrer status

Method sequence

Running sequence

******** counter 2/12
START 03 WAIT 11 s

Parameter

Current command with line number

Editing mode

Menu title

Menu line or command

>Sample sequence
1 MOVE 1 : sample

2

nd

parameter

1

st

parameter