Metrohm 838 Advanced Sample Processor Manual User Manual

Page 70

Advertising
background image

2.8 Dosing and liquid handling

62

Metrohm 838 Advanced Sample Processor, Operation

Exchanging Dosing unit

22

Example:

DOS: 1.2 : release mL

Default port = 2

This command can be used before exchanging a Dosing unit to fill the

cylinder via the given port. In this way, for example, air can be aspi-

rated via Port 4.

After the cylinder has been filled the stopcock will move to Port 2. The

dosing drive can then be removed from the Dosing unit.

Preparing the Dosing unit

Example:

DOS: 1.* : prepar. mL

Default port = 1

In order to prepare a Dosing unit for use a complex procedure is

started: first the cylinder contents are ejected via the given port, then

the volume of the filling tubing is drawn in and ejected again via the

given port. All the connected pieces of tubing are then filled bubble-

free (tubing length > 0 mm) and finally the cylinder is filled again.

The volumes of the connected pieces of tubing (calculated from length

and diameter) are taken into account during the whole procedure.

Automatic emptying

44

Example:

DOS: 1.4 : empty mL

Default port = 1

Automatically emptying the Dosing unit takes place according to a

complex procedure: first the cylinder contents are ejected via the given

port, the ejection tubing is then briefly rinsed with reagent from the fill-

ing tubing and then all the pieces of tubing are emptied in sequence.

Air is always aspirated via the drain port (default: Port 4) for emptying.

The volumes of the connected pieces of tubing (calculated from length

and diameter) are taken into account during the whole procedure.