Special beakers – Metrohm 838 Advanced Sample Processor Manual User Manual

Page 75

Advertising
background image

3.1 Sample racks

Metrohm 838 Advanced Sample Processor, Appendix

67

Special beakers

Special beakers are reserved positions on a sample rack. Up to 16

special beaker positions can be defined per rack. They can be selec-

tively moved to during a run sequence without interrupting the sample

series run or interfering with it. Special beakers can be used for rinsing

the electrode during a sample sequence or for calibrating an electrode

in a start sequence (buffer solutions), etc.

Reserved special beaker positions are automatically recognized in a run

sequence and are skipped over when the individual sample beakers

themselves are being processed.
Special beakers are moved to with

MOVE 1 : spec.1

.

The following settings can be made separately for each special beaker

position on a rack:

Rack position
Working height at Tower 1
Working height at Tower 2
Beaker radius
Beaker sensor


If a special beaker is required in a run sequence, but the Sample Proc-

essor cannot find a beaker at the reserved position, then an error mes-

sage will always appear.