2.4 Commands

38

Metrohm 838 Advanced Sample Processor, Operation

2.4.2 Switching components

Pumps and stirrers can be switched as required, either together or

separately. They work independently and can be operated at the same

time as other functions are being carried out.

PERISTALT

PUMP

4

>sample sequence

4 PERISTALT: on s 3
on,off,1...999

-15…-1, 1…3…15

Pump control

1

st

parameter: status or duration in s

2

nd

parameter: pump rate and direction

With the PERISTALT command the peristaltic pump can be controlled.
The first parameter switches the peristaltic pump on or off. Alternatively

a particular pump duration (in seconds) can be set.
The second parameter sets the pump rate in 15 steps. The signed

value (+ or –) also defines the pump direction. If a negative signed

pump rate is set, the roller head of the pump drive revolves counter-

clockwise, otherwise it revolves in clockwise direction.
The LEARN mode is very useful for determining the optimal pump time

(see p. 71).

PUMP

0

EXT

>sample sequence

4 PUMP 1.1

: 1 s

1.1,1.2

1…999 s,

1.∗,2.∗ on,off

Switching the pump connectors

1

st

parameter: select pump connector

2

nd

parameter: status or duration

With the PUMP command one or both pump connectors (16 V, DC) on

the rear side of the tower can be controlled separately. The first pa-

rameter selects the connector.
Syntax Of the 1

st

parameter: tower.pump

1.* affects both pump connectors (

tower 2 not possible)

The connectors can be switched on or off individually or be activated for

a particular duration.
The LEARN mode is very useful for determining the optimal pump time

(see p. 71).