2.7 Manual operation

50

Metrohm 838 Advanced Sample Processor, Operation

9

LIFT

manual operation
LIFT: 1 : rest mm

work,rinse,shift,

special,rest,

0…235 mm

Position the lift

With the LIFT command, the lift can be run to a given position.

In addition to selecting an exact position in mm (0 – 125 mm)

the <SELECT> key can select a predefined position (work po-

sition, rinse position, shift position, special position, rest position

= 0 mm). The lowest possible lift position can be set in the con-

figuration menu, see section 2.2.2.

2.7.3 Setting the sample position

7

SAMPLE

manual operation
SAMPLE: = 1

1…999

Define current sample

position

The <SAMPLE> command serves to set the current sample posi-

tion. It defines the first sample tube for the subsequent sample se-

ries.


2.7.4 Pump control

PUMP

4

manual operation
PERISTALT: on s 3

on,off

-15…-1,1…15

Switch peristaltic pump

on/off

The PUMP command is used to switch on/off the peristaltic pump at

the 838 Advanced Sample Processor. The PERISTALT command

switches on or off the peristaltic pump, according to the current

state. Pump rate and rotation direction of the pump drive can be set

in 15 steps. A negative value of the pump rate defines a counter-

clockwise rotation. A positive value of the pump rate defines a

clockwise rotation.

2.7.5 Switching the stirrers

5

STIR

manual operation
STIR: T1 : 1 s

T1, T2, T*,
MSB1, MSB2,
MSB 3; MSB*,*

on,

off,

1…9999

s

Stirrer control