2.2

Configuration

Metrohm 838 Advanced Sample Processor, Operation

21

>recall rack

beaker sensor:

off

Selection of the beaker sensor

Tower, SwingH, off

Not implemented.

>recall rack

rack offset

0.00°

Correction for the rotation angle of the

sample rack *

)

-5.00…0.00…5.00°

*

)

All the above lift positions and the rack offset can be set by using the

<LEARN> function.


Special beaker positions

>>recall rack

>>>special positions

Submenu special beaker positions

Open with <ENTER>

Special beaker positions are predefined places on a sample rack that are not

treated as sample positions. They can be occupied by rinsing beakers or

conditioning beakers and selectively addressed during a run sequence. Up

to 16 special beaker positions can be defined per rack. For each special

beaker the work position of the lift and the beaker radius can be set, see

above.


Selection of special beaker

>>>special positions

special beaker

1

1…16

Rack position of the special

beaker

>>special positions

1

rack position

0

etc. up to special beaker 16

0…number Pos.

0 = not defined

>rack definitions

>>store rack

Save rack definitions

In order to store modifications to the definition of a sample rack the submenu

>>store rack

is selected.

>store rack

name: 6.2041.310

Name of rack

10 ASCII characters

With the <SELECT> key you can select one of the existing rack names. The

rack data is stored with <ENTER>. Any rack name can be used. The input

of any rack name can be made directly via the numerical keys or in the text

Use <QUIT>

for next higher