2.8 Dosing and liquid handling

56

Metrohm 838 Advanced Sample Processor, Operation

Dosing command

The following functions can be executed with Dosinos. For each com-

mand the dosing drive and the Dosing unit port at which the command

is to be executed can be entered. The addition of a particular (positive)

volume and filling the cylinder are also possible with 685 and 805 Do-

simats.
If * is entered as wild card for the drive then the selected function will

be carried out by all the connected dosing devices.
If * is entered as wild card for the port then the particular function will

be carried out at the port defined as standard in the parameter menu

under

>def. Dosing units

. This means that this setting is method-

specific, but also applies to manual operation depending on the loaded

method.

Dosing

DOS: X.Y : xxx.xx mL

Adds a particular volume

The given volume will be ejected though the selected port. If a value

with a negative sign is entered then the volume will be aspirated.
The Dosing unit will not be refilled after each addition. The standard

dosing port (Port *) is that defined under
>dosing drive X

dosing port Y
(default: Port 1).

Filling

DOS: X.Y : fill mL

Fills the cylinder

The Dosing unit is filled completely. The liquid is aspirated via the given

port. The standard filling port (Port *) is that defined under
>dosing drive X

filling port Y
(default: Port 2).

Prepare

DOS: X.Y : prepar. mL

Prepares = rinses and fills the

connected tubing and the dos-

ing cylinder

The tubing system of the Dosino should be freed from air bubbles at

least once per day by carrying out a preparation cycle. This is a process

that could take some time.
It is recommended to use this command in a start sequence.
During the preparation process the dosing cylinder as well as the con-

nected tubing are completely filled. Several filling and dosing processes

are carried out. The volumes required for this are calculated from the

configuration settings of the Dosing unit, i.e. from the tubing lengths

and diameters (see Section 2.2.4).