2.6 Run control

48

Metrohm 838 Advanced Sample Processor, Operation

2.6 Run control

START

With <START> you can start a method from the normal operating

condition. If no manual intervention is made, or if no unexpected error

occurs, the sample series will be processed correctly and terminated

with the final sequence. The sample sequence will be repeated several

times in accordance with the entry made under

<PARAM>, number of

samples

, starting with the sample beaker defined as SAMPLE.

If

External START

(see configuration 2.2.1) is switched on then the ac-

tivation of the Remote line Input 7 will also start the method.

STOP

If the sample series is halted with <STOP> then the Sample Proces-

sor will return directly to the normal operating condition. Unprocessed

samples will not be taken into account and the final sequence will not

be carried out. If under Manual stop settings have been defined for

this case, then the corresponding actions or commands will be carried

out, e.g. connected devices will be stopped.
If

external START

(see configuration 2.2.1) is switched on then the ac-

tivation of Remote line Input 6 will also stop the method.

HOLD

LEARN

With <HOLD> the method run can be interrupted. The currently active

command will be terminated immediately. With <START> the method

can be continued with the following command in the active sequence.

The connected peripheral devices will not be interrupted with the

<HOLD> key.

QUIT


<QUIT> breaks off the command which is currently being executed

and starts the next command line in the sequence.
If a fault occurs during a sample series an appropriate error message

will be shown; this must be confirmed with <QUIT>. The Sample

Processor then returns to the HOLD status (see above). After the fault

has been remedied the run can be continued with <START> or, if

necessary, cancelled with <STOP>.

RESET

CLEAR

<CLEAR> interrupts a sample series after the currently active se-

quence has been carried out (gentle stop). The current sample will

be processed to the end.