3.4 Disabling keypad functions

72

Metrohm 838 Advanced Sample Processor, Appendix

• Accept the value with the <ENTER> key (or start the LEARN function

again)

• The LEARN LED goes out; the next command line appears

The LEARN function is available for the following commands:
Command

Adaptive parameter

Type of function

LIFT

Lift position in mm

absolute

PUMP

Pumping time in s

additive

STIR

Stirring time in s

additive

WAIT

Waiting time in s

additive

DOS

Dosing volume in mL

additive

SCN Rm

Status of 8 Remote lines

"live" value

SCN RS

Received character string

"live" value

3.4 Disabling keypad functions

Certain user dialog areas can be made inaccessible to untrained users.

Various dialog areas or keys can be disabled. For example, this can be

used to prevent a method from being overwritten accidentally or pa-

rameters from being altered.
The

>keyboard options

menu for the corresponding settings is opened

by keeping the <CONFIG> key pressed down when the Sample Proc-

essor is switched on. This menu can still be accessed even if the whole

keypad has been disabled.
The individual key functions that can be disabled are:

3.4.1 Disable whole keypad

In routine work, if only a single particular method is used, it may be ad-

visable to prevent manual manipulations on the Sample Processor. For

this reason it is possible to disable (almost) all the keypad keys. The

<START>, <STOP> and <CLEAR/RESET> keys always remain

accessible, so that it is always possible to start and stop methods.

lock keyboard: ON

disables all the keypad keys (with the above ex-

ceptions).

3.4.2 Disable configuration

The configuration of the Sample Processor can be protected against

being overwritten. All the settings in the configuration menu and it's sub-

menus can then no longer be accessed.

lock <CONFIG>: ON

disables the <CONFIG> key.

3.4.3 Disable parameter

If user-defined methods are normally used then it may be advisable to

prevent the stored method parameters from being overwritten. The pa-

rameter menu and it's sub-menus can be made inaccessible.

lock <PARAM>: ON

disables the <PARAM> key.