2.8 Dosing and liquid handling

Metrohm 838 Advanced Sample Processor, Operation

63

Ejecting

11

Example:

DOS: 1.1 : eject mL

Default port = 1

The whole cylinder contents are ejected via the given port. The piston

is lowered beyond the max. volume mark to the stop.

This command should be used for removing air bubbles.

Run piston to end volume

11

Example:

DOS: 1.1 : endVol mL

Default port = 1

The cylinder content is ejected via the given port. The piston is run to

the max. volume mark.

This command should be used for the pipetting functions.

Compensating the mechanical play

44

Example:

DOS: 1.4 : compen. mL

Default port = 1

As the Dosing units are exchangeable, the coupling of the Dosino

connecting rod (spindle) has a slight degree of mechanical tolerance

which is noticeable when the direction of the piston movement

changes. This tolerance can be compensated. This is done by first

making a short piston movement in the same direction as the previous

piston movement, followed by an equivalent piston movement in the

opposite direction.


Stopcock rotation

33

Example:

DOS: 1.3 : port mL

Default port = 1

The stopcock rotates to the given port. No piston movement takes

place. The direction of rotation is determined by the parameter

stop-

cock direction

under

>Dosing unit def.

in the parameter menu.