2.1 Operating principles

Metrohm 838 Advanced Sample Processor, Operation

9

0

EXT

The <EXT> key is used to switch on and off the pump connectors on

the rear side of the tower. On entry of the address of the connector (1

or 2) the current state of the connected pump is switched, i. e. a cur-

rently active pump is switched on.

The display indicates the current state of the pump conncectors (i. e.

PUMP–+

;

+

means active,

means inactive).

Example:

PUMP on/off no. ?

<2> display:

PUMP -+--

PUMP on/off no. ?

<2> display:

PUMP ----

In this case pump 2 is switched on and off.

Under <PARAM>,

>manual stop

you can define whether the pump

connectors should be switched off with the <STOP> key or not.

.

<

VALVE

+

The <VALVE> key is used to switch the position of the injection

valve at the tower of the Advanced Sample Processor.
Example:

VALVE: inject

<ENTER> switches the injection valve to the displayed position.

This function is only working in the Sample Processor models

2.838.0x20 with built-in injection valve.

>

*

RACK

-

With the <RACK> key the sample rack can be initialized. The con-

nected peripheral devices (e.g. Dosimats, Dosinos) are not affected

by this.

The sample rack and lift (both lifts in 2-tower versions) are moved to

the zero position and automatic rack recognition is carried out. The

SAMPLE variable (= rack position of the current sample) is reset to

1.

Sequence control

START

The <START> key starts a method. A start is only possible when the

Advanced Sample Processor is in the normal operating condition, i. e.

when the display shows

ready

.

If <START> is activated after an interruption (<HOLD>, see below),

then the sequence continues with the next command.

The <START> can also be used to carry out a single command line

in a run sequence (TRACE function), see p. Fehler! Textmarke nicht

definiert..