2.8 Dosing and liquid handling

Metrohm 838 Advanced Sample Processor, Operation

55

The Dosing unit has four ports (inlets/outlets) to which various functions

can be assigned. There is an additional outlet (VENT or 0) that is used

for venting the bottle on which the Dosing unit is mounted.

Fig. 7 Dosing unit from below

VENT/0

– is used for venting the storage bottle and is normally fitted with an ab-

sorber tube (filled with a drying agent).

Port 1

– is side-mounted and is defined as dosing outlet 1 as standard.

Port 2

– is located on the base and is defined as the filling inlet as standard; it is

normally fitted with a standpipe.

Port 3

– is side-mounted and is defined as dosing outlet 2 as standard.

Port 4

– is located on the base and is defined as the air inlet opening for empty-

ing the tubing system as standard.

The maximum dosing and filling rates, which can be entered for each

port of the Dosing unit in the configuration menu under

>Dosing units

,

depend on the cylinder size:

Cylinder volume

Max. dosing rate

Resolution

2 mL

6.6 mL/min

0.2 µL

5 mL

16.6 mL/min

0.5 µL

10 mL

33.3 mL/min

1.0 µL

20 mL

66.6 mL/min

2.0 µL

50 mL

160 mL/min

5.0 µL

VENT/0

Port 3

Port 2

Port 1

Port 4