5 auxiliary commands, Wait, Waiting time – Metrohm 838 Advanced Sample Processor Manual User Manual

Page 53: Rack, Initialize rack, Auxiliary commands

Advertising
background image

2.4

Commands

Metrohm 838 Advanced Sample Processor, Operation

45

2.4.5 Auxiliary

commands

WAIT

WAIT

>sample sequence

11 WAIT: pause 1 s
pause, runtime

0…1…9999 s

Waiting time

The WAIT command is used for setting a particular waiting time or for

waiting until a particular time (run time) in a run sequence.
If

pause

is selected then the method run will be interrupted for the en-

tered duration (in seconds).
If

runtime

is selected then the method run will be interrupted until the

selected running time (seconds counter) has been reached. The run-

ning time (in seconds) always starts at the beginning of an individual

sequence, i.e. start sequence, sample sequence or final sequence. If

the running time of the sequence has already been reached or exceed

when the WAIT command is executed then the method will be contin-

ued immediately.

RACK

>

*

RACK

>Sample sequence

12 RACK

Initialize rack

The rack turns to the starting position, i.e. the lift (or both lifts) move

upward to the rest position and the rack is rotated to the initial position.

The rack code is read off and the SAMPLE variable (position of the cur-

rent sample) reset to 1. The RACK command should only be used in a

final sequence.

Advertising