2.7 Manual operation

Metrohm 838 Advanced Sample Processor, Operation

49

2.7 Manual operation

2.7.1 Turning the sample rack / Positioning the samples

NEXT

PREV

With the < NEXT> and < PREV> keys the sample rack

can be rotated forward or backward by one position.

< /NEXT> moves the next sample position (SAMPLE+1) to

the tower. < /PREV> moves the previous sample position

(SAMPLE-1) to the tower.

8

MOVE

Handbetrieb
MOVE 1 : Probe

sample,

ext.1…ext.4

spec.1…16

prev.,

next

+swing, -swing

+rotate, -rotate

1…999

+/-1…999

Position vessel /

Turn rack

With the MOVE command rack positions are moved to, i.e. the

turntable rotates until the sample rack is positioned so that the

selected rack position is placed under the needle. With <SE-

LECT> the numerical rack position as well as the predefined

current sample (SAMPLE command) or the special beakers 1-

16 can be chosen. Furthermore, with the robotic arm external

position (not on the rack) can be accessed. More details about

the MOVE command you can find in section 2.4.1.

For security reasons turning the sample rack is only possible

when the lift is in or above the shift position. The lift is auto-

matically raised, if its current position is lower than the prede-

fined shift position.


2.7.2 Moving the lift

The keys < > and < > allow upward and downward move-

ment of the lift. The lowest possible lift position is defined by the

configuration parameter

'

max. stroke path

'

.

HOME

END

The <HOME> key runs the lift to the rest position (0 mm), i.e. to

the upper limit.
<END> runs the lift to the predefined work position (see sec-

tion 2.2.3).

Advertising