2.4 Commands

42

Metrohm 838 Advanced Sample Processor, Operation

1

SCAN

>Sample sequence

8 SCN:RS

Rm,RS

default value: ∗R"

14 ASCII characters

Scanning the RS232 interface

= scan "Ready" status message

any data string with 14 characters

In a sequence the SCN:RS command will halt the method sequence

until the predefined data string (up to 14 characters) has been received

via the serial RS232 interface. Incoming data is checked character by

character.
Make sure that the transmission parameters of the RS232 interface are

identical with those of the connected device (see configuration menu

>RS232 settings

, p. 23).

Any letters, numbers and special characters can be selected from the
character set of the Sample Processor. '∗' can be set as wild card for
any character string. (If '∗' is to be interpreted as an ASCII character
then '∗∗' must be used.) A wild card can be set within a character string.

When the first part of the character string has been recognized correctly

then a search will be made for the first appearance of the character
standing after '∗'. In this case the comparison will be made with the

second part of the character string.
Details of the CTL command are given below.
With the LEARN function transmitted data (=character strings) can be

adopted interactively (see p. 71).