3.6 Error messages

76

Metrohm 838 Advanced Sample Processor, Appendix

3.6 Error messages

QUIT

If a fault should occur then the execution of the active command will be

stopped and an error message shown (display blinks). This must be

confirmed with the <QUIT> key.

If the Sample Processor is processing a series of samples when the

fault occurs then it will switch to the HOLD status. When the cause of

the fault has been found and remedied the sample series can be re-

started with the next command by pressing the <START> key. If it is

not possible to remedy the fault then the method can be stopped with

<STOP>.

List of possible error messages and their causes:

battery low

The battery for the permanent storage of the user data must

be replaced.

beaker too small

The rack position which has been moved to has a smaller

beaker radius than that defined as the min. beaker radius for

the particular lift. There is a risk of damage. Check the sam-

ple rack, its configuration settings and those of the selected

tower.

changer overload

Load too high, or resistance to the action to be carried out.

After <QUIT> the Sample Processor will be initialized again.

dos.## not exec

A fault has occurred at the given dosing device.

dos.## not ready

The selected dosing device cannot carry out the particular

command as it is carrying out a different action or the current

instrument status does not allow it to be carried out.

dos.unit ## missing

The selected dosing device has not been connected.

dos.unit ## overload

The selected Dosing unit cannot carry out a dosing com-

mand. Check the buret and piston.

invalid position

The selected sample position does not exist or is defined as

a special beaker or the selected special beaker has not been

defined. The selected lift position may lie outside the max.

stroke path.

invalid rack code

The rack code read in by the Sample Processor cannot be

found in the internal table.

rack data missing

No sample rack in position, or no rack data can be found for

the sample rack currently on the turntable.

RS232 error

The transmission parameters of the RS232 interface do not

coincide with those of the receiving device.