2 connection to ic system – Metrohm 750 Autosampler User Manual

Page 13

Advertising
background image

2 Installation

750 Autosampler

10

2.2

Connection to IC system

2.2.1

750 Autosampler as "Master"

When an IC system consisting of 732 IC Detector, 733 IC Separation

Center, and 709 IC Pump is operated with the 750 Autosampler as

"Master" the command to start the next injection after the analysis time

"

Run Time

" has elapsed will be given by the Autosampler itself. In this

case the 750 Autosampler is connected to the 733 IC Separation Cen-

ter with the 6.2128.100 cable as shown in Fig. 6. In addition the 709 IC

Pump can be connected to the 750 Autosampler in such a way that

when the pump stops the Autosampler will also be automatically

stopped (important for unattended continuous operation).

1 Electrical connection 750 – 733

Connect 6.2140.020 connector plug to contact closure strip 99

of the 750 Autosampler.

Connect one end of the 6.2128.100 cable to the connections

"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM"

on the connector plug inserted in the contact closure strip 99.

The four cable ends are appropriately inscribed with "LOAD

NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and

"INJECT COM" (white).

Connect the other end of the 6.2128.100 cable to the con-

nections "Ground", "Fill" and "Inject" on the terminal block 38

38

(valve A) or 46

46 (valve B) of the 733 IC Separation Center. The

three cable ends are appropriately inscribed with "GND"

(white), "FILL" (green) and "INJECT" (brown).

2 Electrical connection 750 – 709 (optional)

Connect 6.2140.010 connector plug to contact closure strip

10

10 of the 750 Autosampler.

Connect jumper P12 to the control interface of the 709 IC

Pump in such a way that the contacts at outputs 3 and 4 will

be closed when the pump has stopped (see chap. 5.2 of the

709 Instructions for Use).

Use 6.2134.070 cable (option) to connect outputs 3 and 4 of

the control interface on the 709 IC Pump with connection

"REMOTE STOP INPUT" and "GND" on contact closure strip

10

10 of the 750 Autosampler.

Settings with 750 Autosampler as "Master"

The following settings are recommended for operating the IC system

with 750 Autosampler as "Master":

• 750 Autosampler:

"

Run Time

"

3.0 min

• 732 IC Detector:

operation without program or

with "Inject" program