8 remote control – Metrohm 750 Autosampler User Manual

Page 39

Advertising
background image

3 Operation

750 Autosampler

36

3.8

Remote control

REM/LOC

6

By pressing the <REM/LOC> key (key <FUNCTION> + <6>) the

menu for switching between normal operation (Local) and remote-

controlled operation via the RS232 interface (Remote) appears. The key

opens the following menu:

REMOTE/LOCAL
Local
RS-232
IEEE-488
Exit

Remote control/Normal operation

Normal operation

Remote control via RS232

Remote control via IEEE-488

Return to standby mode

The individual menu points are selected by pressing the <

б

б

> and

<

в

в

> keys. The individual inquiries of a menu point are accessed using

the <ENTER> key and exited by pressing simultaneously the

<FUNCTION> + <CLEAR> keys. The following listings show all dia-

log items which appear under <REM/LOC>. The values shown in the

displays are the default values, the possible entry values or ranges are

shown below the display.

REMOTE/LOCAL

à

Local

Switch to normal operation
After confirmation of this command with <ENTER> the instrument

is switched from operating under remote control via RS232 to

normal operation.

REMOTE/LOCAL

à

RS-232

Switch to remote control via RS232
After confirmation of this command with <ENTER> the following

display appears:

Select Baud Rate

<4800>

300,600,1200,2400,

4800,9600,19200,

38400

Entry of baud rate
Entry of baud rate in bits/s for the RS232 interface. The value

entered here must also be set on the external controller. After

the baud rate has been confirmed with <ENTER> the fol-

lowing display appears:

Select Parity

<None

Even,Odd,None

Entry of parity
Entry of parity for the RS232 interface. The value entered

here must also be set on the external controller. After the

parity has been confirmed with <ENTER> the following

display appears: