5.2 Interfaces

750 Autosampler

51

5.2

Interfaces

5.2.1

Contact closure strips

The contact closure strip 99 can be used
for external control of an injection valve.
The allocation of the 4 connections is
shown in Fig. 12. The output signals are
emitted as contact closures between two
connections.

LOAD (output signal for switching the
injection valve)
Automatic contact closure between connections 1 and 2 (normally
open) before the start of the filling process. This signal is used to switch
the injection valve of the 733 IC Separation Center to the "Fill" position.

INJECT (output signal for switching the injection valve)
Automatic contact closure between connections 3 and 4 (normally
open) at the end of the filling procedure. This signal is used to switch
the injection valve of the 733 IC Separation Center to the "Inject" posi-
tion.

The contact closure strip 10

10 can be used

both for external control of the 750 Auto-
sampler as well as for the control of ex-
ternal instruments via the 750 Autosam-
pler. The allocation of the 19 connections
is shown in Fig. 13. The output signals
are emitted as contact closures between
two connections; the input signals must
be applied in the same way between two
connections.

START (output signal for injection)
Automatic contact closure between con-
nections 1 and 2 (normally open) and 1
and 3 (normally closed) for each injec-
tion triggered by a method. Under
"

SYSTEM SETTINGS/Contact Closures/In

ject

" the contact closure time can be set

between 1

32000 s (for manual injections the contact closure time is

always 6 s). This signal can be used, for example, for starting the data
evaluation system.

T1 (output signal for freely programmable event)
Contact closure between connections 4 and 5 (normally open) and 4
and 6 (normally closed) according to the time laid down in the method
"

Event1 Time

". The contact closure time can be set under "

SYSTEM

SETTINGS /Contact

Closures/Event

1

" between 1

32000 s. This signal

can be used for controlling external instruments.

REMOTE INJ. VALVE

LOAD

8
7
6
5
4
3
2
1

INJECT

Fig. 12: Valve control

NO CONNECTION

READY INPUT

REMOTE STOP INPUT

GND

+5 VOLTS OUTPUT

GND

EXTERNAL INJECT INPUT

NC

NO

COM

NC

NO

COM

NC

NO

COM

NC

NO

COM

{

START

19
18
17
16
15
14
13
12
11
10

9
8
7
6
5
4
3
2
1

T1

T2

T3

Fig. 13: Inputs/outputs