2.2 Connection to IC system
750 Autosampler
13
The four cable ends are appropriately inscribed with "LOAD
NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and
"INJECT COM" (white).
•
Connect the other end of the 6.2128.100 cable to the con-
nections "Ground", "Fill" and "Inject" on the terminal block 38
38
(valve A) or 46
46 (valve B) of the 733 IC Separation Center. The
three cable ends are appropriately inscribed with "GND"
(white), "FILL" (green) and "INJECT" (brown).
732
733
709
762
PC
6.2134.100
6.2128.130
6.2115.070
6.2134.090
6.2125.090
(732)
INJECT
LOAD
750
6.2134.000
6.2128.100
Fig. 7:
Connection of 750 Autosampler at 732/733 and 762