2.3 Mains connection

750 Autosampler

15

2.3

Mains connection

Follow the instructions below for connecting to the power supply. If

the instrument is operated with the mains voltage set wrongly and/or

wrong mains fuse there is a danger of fire!

2.3.1

Check fuses

The selection of mains voltage (100, 120, 220 or 240 V) and mains fre-

quency (50 or 60 Hz) is made automatically by the 750 Autosampler

according to the mains voltage and frequency used. Nevertheless, be-

fore the 750 Autosampler is switched on for the first time a check

should be made as to whether the correct fuses for the intended mains

supply have been inserted in the instrument. Proceed as follows:

1 Disconnect mains cable

Disconnect mains cable from mains connection plug 14

14 of the

750 Autosampler (see Fig. 5).

2 Remove fuse cassette

Using a screwdriver, loosen fuse cassette 15

15 alongside the

mains connection plug 14

14 and take out completely.

3 Remove fuse holder

Lift up the spring clip on the upper side of fuse cassette 15

15 and

remove the fuse holder completely from the fuse cassette.

4 Check fuses

Carefully take the two fuses out of the fuse holder and check

their specifications:

100

…120 V

2 A

(rapid-action, with low rup-

turing capacity, 3AG)

230

…240 V 1.25 A (rapid-action, with low rup-

turing capacity, 5

××20 mm)

5 Insert fuses

Change fuses if necessary and reinsert in fuse holder.

6 Install fuse holder

Push the fuse holder into fuse cassette 15

15 until the spring clip

clicks into position.

7 Insert fuse cassette

Completely reinsert fuse cassette 15

15 in 750 Autosampler.