2.4 Initial checkout

750 Autosampler

17

2.4

Initial checkout

Initial checkout will prove the general operation of the instrument and

show that no shipping damage or installation problem exists which

would prevent normal operation. To carry out this checkout the 750

Autosampler must be connected to the 733 IC Separation Center as

described in chap. 2.2 and correctly connected to the mains supply as

described in chap. 2.3.
For the initial checkout a standard chromatogram for the separation

column used in the 733 IC Separation Center should be recorded. In

order to do this a sample vial 44 filled with at least 500

µ

L of a suitable

standard ion solution should be placed in position 1 of sample rack 33;

then proceed step by step as follows:

1 Switch on 750 Autosampler

Switch on 750 Autosampler with mains switch 13

13. The following

screen will be displayed after powerup and initialization:

READY

SELECT A FUNCTION

2 Select "MANUAL INJECTION"

Press the <MANUAL INJECTION> key (<FUNCTION> +

<8>). The following screen will appear:

MANUAL INJECTION

а

а

Sample Vial 1

Needle Depth POS 1
Run Injection

в

в

3 Enter sample vial position

Press the <1> key and then press the <ENTER> key. The

screen will now change to:

MANUAL INJECTION
Sample Vial 1

а

а

Needle Depth POS 1

Run Injection

в

в

4 Enter needle depth

Press the <

> key two times and then press the <ENTER>

key. This results in the following display:

MANUAL INJECTION
Sample Vial 1
Needle Depth POS 3

а

а

Run Injection

в

в