3.2 System settings

750 Autosampler

25

NEEDLE SETUP

à

Pressure Low>

Low,High

Pressure for needle lowering
The needle can be lowered either with "

Low

" pressure or with

"

High

" pressure.

If sample vessels closed with 6.2743.030 filter stoppers are used then

pressure "

High

" must be selected.

SYSTEM SETTINGS

à

Key beep

On/off switching of an audible beep
By pressing <ENTER> the following screen is displayed:

Audible Key Feedback
Selection

<ON

ON,OFF

Audible beep whenever a key is pressed
Switches acoustic signal each key a key is pressed on/off.

Switching is made by means of the <

а

а

> or <

Я

Я

> cursor

keys.

SYSTEM SETTINGS

à

Time/Date

Date/time settings
By pressing <ENTER> the following screens are displayed:

Time/Date

à

02/23/00 03:31 AM

Display of date and time

Time/Date

à

Set Date

Set date
By pressing <ENTER> a menu is opened for setting "

Year

",

"

Month

", and "

Day

".

Time/Date

à

Set Time

Set time
By pressing <ENTER> a menu is opened for setting "

Hour

",

"

Month

", and "

AM

" or "

PM

".

SYSTEM SETTINGS

à

Injection Valve

Injection valve
Without function for 750 Autosampler.

SYSTEM SETTINGS

à

Status

Status display
By pressing <ENTER> the following screen is displayed:

Validated
709/50 S/N 03029
Version H1.00A1.00
<CLEAR> to Exit

Status display
Display of validation status, instrument type, serial number,

and program version number.