5.2 Interfaces

750 Autosampler

53

5.2.2

RS232 interface

The RS232 interface 11

11 is used for complete control of the 750 Auto-

sampler by an external computer. The interface is wired as DCE device
(DCE: Data Communication Equipment) and has the following technical
data:

Baud rate: 300

38400 (selectable, see section 3.8)

Data bits: 5

8 (selectable, see section 3.8)

Parity:

even, odd, none (selectable, see section 3.8)

Stop bit:

1, 2 (selectable, see section 3.8)

A 1:1 RS cable with a male 9-pole plug is required as a connection ca-
ble (e.g. 6.2134.000 Metrohm cable 9-pol/9-pol).
Remote operation of the Autosampler is carried out with the ASCII
commands described in chap. 3.8. An example of the use of these
commands in a RS232 control program, which is programmed in Mi-
crosoft Basic, is given below.

100 ' *** OPEN COM1 FOR RS-232 ***

110 OPEN "COM1:4800,N,8,1,RS,CS,DS,CD" AS #1

120 ' *** SEND CARRIAGE RETURN TO COM1 **

130 PRINT #1,CHR$(13)

140 ' *** PUT VALVE INTO THE LOAD POSITION ***

150 PRINT #1,"@VLV LD" + CHR$(13)

160 ' *** GO TO NEEDLE HOME POSITION ***

170 PRINT #1,"@GTV 0" + CHR$(13)

180 ' *** GO TO VIAL POSITION 25 ***

190 PRINT #1,"@GTV 25" + CHR$(13)

200 ' *** LOWER THE NEEDLE TO POSITION 1 ***

210 PRINT #1, "@LDL 1" + CHR$(13)

220 '*** PUT VALVE IN INJECT POSITION ***

230 PRINT #1, "@VLV INJ" + CHR$(13)

240 ' *** CLOSE RS-232 PORT ***

250 CLOSE #1

260 END