4.2 Maintenance and servicing

750 Autosampler

41

4.2

Maintenance and servicing

4.2.1

General information

The 750 Autosampler was designed for years of maintenance-free use.

The X,Y,Z sample tray assembly is equipped with bushings and bear-

ings made from materials chosen to minimize the need for lubrication.

The needle drive carriage position is recalibrated at powerup and with

the start of each method. Therefore, there are no user serviceable parts

in the X,Y,Z sample tray assembly or the needle drive carriage with the

exception of changing the transfer needle (see section 4.2.2).

4.2.2

Transfer needle replacement

If the transfer needle is blocked it must be replaced. Proceed as follows

(see Fig. 11):

1

Prepare Autosampler

Stop any method which may be running by pressing

<RUN/STOP> and <ENTER>.

Park the transfer needle 55 by pressing <SYSTEM> and "

Park

Needle

".

Switch off instrument by means of mains switch 13

13.

2

Remove old needle

Hold connection screw 21

21 with open-end spanner and un-

screw needle holder 17

17 together with transfer needle 55 with

open-end

7

/

32

"

spanner (6.2621.020).

Use two open-end spanners to unscrew nut 19

19 from connec-

tion screw 21

21 and remove connection screw 21

21 from holder

20

20.

Remove old transfer needle 55 from needle holder 17

17.

3

Insert new needle

Insert 6.2620.130 spare needle with already mounted ferrule

16

16 in needle holder 17

17.

Insert connection screw 21

21 into holder 20

20 from below.

Use an open-end spanner to screw nut 19

19 firmly onto con-

nection screw 21

21.

Screw needle holder 17

17 onto connection screw 21

21 by hand.

Hold connecting screw 21

21 with open-end

7

/

16

"

spanner,

tighten up needle holder 17

17 with open-end

7

/

32

"

spanner

(6.2621.020) by about half a turn.