3 Operation

750 Autosampler

30

METHOD N STEP M

à

Run Time 3.0

0.0,0.1

999.9 min

Run time
Selects the run time (chromatographic analysis

time) between switching the injection valve to the

"INJECT" position and the start of a new injection

with the same or the next sample.
The minimum run time is 0.1 min. If 0.0 min is

entered the external injection start is activated.

The sample loop will only be filled and the injec-

tion started when a contact closure has been

made at the "EXTERNAL INJECT INPUT". The

run time will then be determined externally.

METHOD N STEP M

à

Event1 Time 0.0

0.0

999.9 min

Time for event T1
Selects the time after which a signal in the form

of a contact closure will be produced at output

"T1".
The signal will only be produced if the selected

time is shorter or the same as the run time. If the

production of a signal is to be avoided for a

sample series then a time must be selected

which is longer than the run time.

METHOD N STEP M

à

Event2 Time 0.0

0.0

999.9 min

Time for event T2
Selects the time after which a signal in the form

of a contact closure will be produced at output

"T2".
The signal will only be produced if the selected

time is shorter or the same as the run time. If the

production of a signal is to be avoided for a

sample series then a time must be selected

which is longer than the run time.

METHOD N STEP M

à

Event3 Time 0.0

0.0

999.9 min

Time for event T3
Selects the time after which a signal in the form

of a contact closure will be produced at output

"T3".
The signal will only be produced if the selected

time is shorter or the same as the run time. If the

production of a signal is to be avoided for a

sample series then a time must be selected

which is longer than the run time.

METHOD N STEP M

à

Rinse Type NONE>

NONE,EVEN,ODD,

QR_E,QR_O

Rinse type
This parameter allows every second sample vial

to be programmed as a rinsing vial. The follow-

ing options can be selected with the <

а

а

> and

<

Я

Я

> cursor keys:

NONE

No rinses. The vials are used as pro-

grammed.