Metrohm 750 Autosampler User Manual

Page 47

Advertising
background image

4 Notes – Maintenance – Faults

750 Autosampler

44

first injection followed by smaller but equal peak heights for the second

two injections are observed. Observation of either pattern requires

complete replacement of the loop.
The correct way to mount seals is described in detail in chap. 2.5 of the

732/733 Instructions for Use.
When using a previously swaged steel capillary, be sure to use it with

the fitting in which its ferrules were originally set. Small inconsistencies

in the machining of the fitting are a source of dead volume.

Submerged waste tube

A waste tube that has been submerged below the liquid level in the

waste container can cause the waste solution to siphon into the sample

loop. The type of non-reproducibility this causes exhibits random peak

heights. Do not allow the waste tube to be submerged below the liquid

level in the waste container.

Injection valve leak

The probable cause of injection valve leakage on the 733 IC Separation

Center is a worn or damaged seal. In this case new seals must be fit-

ted.

4.3.2

Missed injections

Skipped vials

The torque of the stepper motor which drives the transfer needle is lim-

ited so as to prevent the vial from bursting under pressure in the event

that a plug occurs in the transfer needle, transfer tubing, or injection

valve. If more pressure is required to push a cap in than is provided, the

needle motor will stall. While a method is running, if the motor stalls on

two consecutive vials, the Autosampler will stop the method and issue

an error message.
However, if the motor stalling is intermittent, the Autosampler will con-

tinue with the method. A list of the skipped vials will be shown on the

display once the method is complete.
Plugs in the plumbing between Autosampler and injection valve are the

major source for motor stalls and skipped vials (see section 4.3.1).

However, the same symptom can occur with highly viscous samples or

when FilterCaps

are used.

FilterCaps

due to job they are ask to do can pose some minor plug-

ging problems. In some instances, the sample may be too "dirty" for

use with FilterCaps

. A suggested work around for the problem of us-

ing FilterCaps

with really "dirty" samples is to limit the number of injec-

tions per vial to 1 and to set the Needle Depth to "

POS1

" or "

POS2

" using

the system settings key. Should this latter suggestion fail to provide

satisfactory results, an alternative sample filtering procedure should be

used.

Advertising