4 Notes – Maintenance – Faults

750 Autosampler

40

A further possible way of filtering samples is by means of the optional

6.2743.030 filter stoppers. These filter stoppers, which contain a

built-in filter unit, are fitted to the sample vials just like normal stoppers.

The way they work is shown in Fig. 10.

Fig. 10:

Sample filtering with FilterCaps

FilterCaps

were designed to free the chromatographer from the time-

consuming task of hand filtering samples. They can be used to free

samples from suspensions and particles. However, filter stoppers are

not suitable for filtering samples containing a large amount of solids as

this causes the filter to become blocked which also blocks the transfer

needle. Those samples which are difficult to filter manually with a mem-

brane filter (see Fig. 9) should be centrifuged before filtration.

When working with filter stoppers the "

POS4

" setting for the needle

immersion depth "

Needle Depth

" must not be selected.

4.1.3

Operating parameters

Before samples can be injected all necessary instrument settings must

be made. The most important parameter is the retention time. The time

interval "

Run Time

" between individual injections must be sufficiently

long so that all substances are eluted before the next sample is in-

jected. Moreover, it may be necessary to include further time for base-

line analysis and evaluation with the evaluation system.

Filtered sample

Unfiltered sample

To injection valve