4 Notes – Maintenance – Faults

750 Autosampler

48

4.4.3

Remote control error codes

If the remote control for the 750 Autosampler is switched on via the

RS232 interface then each error will produce a corresponding error

code. For each of the errors listed below information is provided as to

whether the error can be remedied (R) or if it is a fatal system error (F).

The Autosampler remains ready for operation with R-errors and the op-

erator can enter a new command. With F-errors it is no longer possible

to use the Autosampler. The instrument must be switched off and re-

paired (see chap. 4.4.2).

Code Error

Type

0

No error, autosampler is not busy and is ready for a

command

.

-1

Invalid auxiliary output contact number request

.

R

-2

Invalid auxiliary output command request

.

R

-4

Request invalid because autosampler is

busy performing an operation

.

R

-5

Invalid buzzer request

.

R

-6

Invalid command request. Autosampler did

not recognize command

.

R

-8

Invalid display request

.

R

-12

Present needle position is unknown or invalid

.

R

-13

Requested needle position is invalid

.

R

-21

Invalid vial number request

.

R

-22

Transfer needle is not at a valid vial position, so it

cannot be lowered

.

R

-27

Invalid injection valve position request

.

R

-28

Motor drive error on x drive (left-right).

F

-29

Motor drive error on y drive (forward-back).

F

-30

Unable to determine vial position or vial position is

invalid

.

R/F

-31

Motor drive error on z drive (up-down).

F

-32

Motor stall error on z drive

.

F

-35

Error, origin unknown

.

F

-36

Autosampler stopped locally

.

R

-43

Step position requested by MVZ not valid

.

-47

Step position requested by MVX or MVY not valid

.

-48

Not a valid X/Y step position

.

-49

Serial communications error

.