3 Operation

750 Autosampler

24

CLOSURE DURATIONS

à

Injection 1

1

32000 s

Contact closure time for "START"
Input of contact closure time for the signal to be produced at

output "START" when the injection valve is switched to the

"INJECT" position.

CLOSURE DURATIONS

à

Event # 1

1

32000 s

Contact closure time for "T1", "T2" or "T3"
Entry of contact closure time for the signal to be produced at

output "T#". The signal output is made according to the time

laid down in the method "

Event# Time

".

CLOSURE DURATIONS

à

Exit

Return to "SYSTEM SETTINGS" menu

SYSTEM SETTINGS

à

Loop Size

Loop size
Size of the loop at the injection valve at the 733 IC Separation

Center. By pressing <ENTER> the following screen is displayed:

Enter Loop Size
10

µµL

(Four injs/vial)

1

100

µ

L

Entry of loop size
Entry of loop size in

µ

L. In the bottom line the number of

possible injections per sample is displayed. For a volume of

10

µ

L the maximum is 4, for a volume > 10

µ

L a maximum

of 3 injections can be made.

SYSTEM SETTINGS

à

Needle Setup

Needle settings
By pressing <ENTER> the following menu is opened:

Depth

Needle depth.

Pressure

Pressure for needle lowering.

NEEDLE SETUP

à

Depth

Needle depth
By pressing <ENTER> the following screen is displayed:

Enter Needle
position for ONE
injection per vial.
<POS 3>

POS1,POS2,POS3,POS4

Entry of needle depth
Entry of the position to which the transfer needle is lowered

with 1 injection per sample. The positions correspond to the

following displacement volumes (see section 1.2):

POS1

310

µ

L

POS3

590

µ

L

POS2

450

µ

L

POS4

730

µ

L

If sample vessels closed with 6.2743.030 filter stoppers are used then

immersion depth "

POS4

" must not be selected.