background image

4.1 Operating information

750 Autosampler

39

4

Notes – Maintenance –

Faults

4.1

Operating information

4.1.1

Loading sample vials

Load the vial with sample to within

6 mm (1/4 in.) of the top and push

the polyethylene cap (cup side down) into the vial. The cap has two

sealing rings on its periphery. The cap should be pushed into the vial

so that both rings engage and remain so. The maximum volume of

sample is 800

µ

L, the minimum volume of sample is 300

µ

L using a

loop size > 20

µ

L and 150

µ

L using a loop size

20

µ

L.

Air trapped under the plastic caps is of no consequence. The first

stroke of the needle into the vial displaces enough liquid to force the air

completely through the injector valve loop and through the waste outlet

tube. Should any air remain in the vial, it will be above the needle

opening and will not enter the tubing. The first stroke also forces

enough sample through the tubing and loop to remove remnants of the

previous sample.

4.1.2

Sample preparation

It is important that samples be free of particulate matter which may plug

the needle or tubing. Metallic or gritty particles will cause damage to the

injection valve and may plug the chromatographic system. All samples

should be filtered therefore prior to filling the vials. This can be accom-

plished using a hand filtration device with a 0.45

µµm membrane filter

shown in Fig. 9.

Fig. 9:

Manual sample filtering