2 732 ic detector as "master – Metrohm 750 Autosampler User Manual

Page 14

Advertising
background image

2.2 Connection to IC system

750 Autosampler

11

732

733

709

6.2125.090

INJECT

LOAD

750

6.2134.070

6.2128.100

6.2125.060

6.2115.070

Fig. 6:

Connection of 750 Autosampler at 732/733

2.2.2

732 IC Detector as "Master"

When an IC system consisting of 732 IC Detector, 733 IC Separation

Center, and 709 IC Pump is operated with the 732 IC Detector as

"Master" the 750 Autosampler sample change procedure will be started

by a signal from the 733 IC Separation Center. The Autosampler must

be connected to the 733 IC Separation Center with two cables as

shown in Fig. 6. In addition the 709 IC Pump can be connected to the

750 Autosampler in such a way that when the pump stops the Auto-

sampler will also be automatically stopped (important for unattended

continuous operation).

1 Electrical connection 750 – 733

Connect 6.2140.020 connector plug to contact closure strip 99

and 6.2140.010 connector plug to contact closure strip 10

10of

the 750 Autosampler.

Connect one end of the 6.2128.100 cable to the connections

"LOAD NO", "LOAD COM", "INJECT NO" and "INJECT COM"

on the connector plug inserted in the contact closure strip 99.

The four cable ends are appropriately inscribed with "LOAD

NO" (green), "LOAD COM" (yellow), "INJECT NO" (brown) and

"INJECT COM" (white).

Connect the other end of the 6.2128.100 cable to the con-

nections "Ground", "Fill" and "Inject" on the terminal block 38

38

(valve A) or 46

46 (valve B) of the 733 IC Separation Center. The

three cable ends are appropriately inscribed with "GND"

(white), "FILL" (green) and "INJECT" (brown).

Evaluation system

Cable 732/Evaluation
system (start signal)

Cable 732/Evaluation
System (analog signal)