Maintenance and Troubleshooting 5-17

Problem

Ultrasonic depth readings are erratic.

Possible

Causes

Noise may exist in the sewer pipe.

Ultrasonic depth sensor may be dirty.

Flow may be choppy or foamy.

Ultrasonic depth sensor may not be level.

Pulse command and spare 2 delay parameters may be
set too low for the pipe.

Possible

Solutions

Check the flow conditions.

Clean the ultrasonic depth sensor.

Check the level of the ultrasonic depth sensor.

Review and adjust the pulse command and spare 2
delay parameters.

Problem

Less than six good pairs are received when firing the
ultrasonic depth sensor.

Possible

Causes

Ultrasonic depth sensor may be dirty.

Ultrasonic depth sensor may be faulty.

Depth board may be faulty.

Possible

Solutions

Clean the ultrasonic sensor.

For less than four good sensor pairs, contact a trained
ADS technician.

Replace the ultrasonic depth sensor.