Switch and Jumper Settings B-3

The following figure shows the correct jumper settings for the 4000
flow monitor backup battery. If the technician disables a board and
places it in storage, the technician must re-configure the backup
battery prior to use.

Storing a monitor requires disabling the backup battery. To disable
the backup battery, locate Jumper 5 (three-position jumper header)
and move the jumper to cover positions 2 and 3. This disables the
battery. Placing the jumper covering positions 1 and 2 initiates
backup battery operation. Use the following table to disable and
enable a backup battery.

The options switch, identified by SW3 on the processor board,
indicates the specific configuration of the monitor, including
parameters such as the method of communication (modem/serial)
and baud rate. For the 4000 monitor, switch 4 is the only switch
that should be in the closed position on the options switch. All
other switches
should remain in the open position. For the 4000
WR
monitor, switches 4 through 8 should be in the closed position
and switches 1 through 3 should be in the open position.

Backup battery configuration and jumper position

J5

1

2

3

J5

1

2

3

J5

1

2

3

Backup Battery
Operating

Battery in
Storage

Three-position
Jumper Header
without Jumper

Options Switch Settings