Maintenance and Troubleshooting 5-19

Problem

Monitor often provides a velocity reading of 0.

Possible

Causes

Doppler velocity sensor may be dirty or broken.

Cabling between the velocity board and the velocity
sensor may be bad.

Velocity board may be faulty.

Velocity parameters may require adjustment.

Possible

Solutions

Clean the velocity sensor.

Check the velocity parameters. Consider increasing the
maximum carrier parameter.

Inspect the velocity sensor cables for tightness.

Check for moisture in the connector.

Contact a trained ADS technician.

Replace the velocity sensor.

Replace the velocity board.

Problem

Velocity data does not fluctuate much (but is not 0).

Possible

Cause

Doppler velocity sensor may be dirty

Velocity sensor may be broken.

Possible

Solutions

Clean the velocity sensor.

Check the velocity sensor connections.

Check the velocity parameters.

Contact a trained ADS technician.

Replace the velocity sensor.

Replace the velocity board.