Pressure depth subsystem – ADS Environmental Services Model 4000 530017 A0 User Manual

Page 106

Advertising
background image

5-20

ADS Model 4000 Manual

Problem

Velocity data seems erratic.

Possible

Causes

Velocity sensor may be dirty.

Velocity sensor may be faulty.

Velocity board may be faulty.

Possible

Solutions

Clean the velocity sensor.

Check the sensor connections.

Check the velocity parameters.

Contact a trained ADS technician.

Replace the velocity sensor.

Replace the velocity board.

Problem

Velocity readings are abnormally high.

Possible

Causes

Wires on the P2 connector may be loose.

Doppler velocity sensor may be covered with silt.

Velocity sensor may be out of the flow.

Flow may be reversed or slower than 0.5 ft/s (0.15 m/s).

Possible

Solutions

Check the site conditions, and relocate the velocity
sensor if necessary.

Replace the velocity board.

Pressure Depth Subsystem

The following tables contain general troubleshooting techniques for
the pressure depth subsystem.

Warning: Contact a trained ADS technician for further
diagnosis before replacing a sensor.

Advertising