Ultrasonic depth subsystem – ADS Environmental Services Model 4000 530017 A0 User Manual

Page 99

Advertising
background image

Maintenance and Troubleshooting 5-13

Problem

Data is missing at the beginning or end of the date range
following data collection.

Possible

Causes

Monitor activation may have failed.

Monitor time may be incorrect.

Monitor’s basic code or variable file may be corrupt.

Possible

Solutions

Verify whether the monitor has been activated, and
activate if necessary.

Check monitor time, and reset clock if necessary.

Run diagnostics in FieldScan to verify whether a basic
code problem may exist.

Contact your regional ADS representative.

Problem

An I/O error message displays when communicating
with the monitor.

Possible

Cause

Processor board, depth board, or velocity board may be
faulty.

Possible

Solutions

Re-attempt communication with the monitor.

Contact your regional ADS representative.

Ultrasonic Depth Subsystem

The following tables contain general troubleshooting techniques for
problems with the ultrasonic depth subsystem.

Note: Contact a trained ADS technician for further
diagnosis prior to replacing a sensor.