Maintenance and Troubleshooting 5-21

Problem

Temperature reading from the pressure depth sensor
shows a value of -273

° C (-459° F).

Possible

Causes

Cable connecting the pressure sensor to the monitor
may be loose or broken.

Pressure sensor may be faulty.

Depth board is not responding or may be faulty.

Possible

Solutions

Secure pressure sensor connection to monitor (if
necessary).

Replace the pressure sensor.

Replace the depth board.

Problem

Pressure depth readings are consistently incorrect.

Possible

Causes

Coefficients may be incorrect.

Pressure dryer tube may not be functioning properly.

Possible

Solutions

Retrieve or enter the correct coefficients, and reactivate
the monitor.

Inspect and replace (if necessary) desiccant in pressure
dryer tube.

Problem

Pressure depth readings are consistently off by up to 3
inches (7.62 cm), but temperature readings are
accurate.

Possible

Cause

Monitor may be configured with an incorrect pressure
sensor offset.

Possible

Solution

Enter an offset to depth for consistency with confirmation
readings.