26 MSI-3460 Operator’s Manual

5.0

Calibration

The MSI-3460 is calibrated using standard weights. It is required that the weight used

is at least 10% of full capacity in order to achieve rated accuracy. For example, use at

least a 500 kg test weight to calibrate a 5000 kg capacity scale. Although a single span

point is usually adequate for rated accuracy, the MSI-3460 supports Multi-Point

calibration with up to 4 span points + Zero.
When adequate test weights are not available, the MSI-3460 can be calibrated using a

calculated constant calibration which is referred to as C-Cal. To use C-Cal, a

previously generated C-Cal number must be known. MSI supplies replacement load

cells for the MSI-3460 with the C-Cal value stamped on the serial number label.
There are three types of calibration:

• Standard Calibration – is used for maintenance and routine calibration.
Initial Calibration – is used to set up both the capacity and resolution (d) of the

scale. It differs from standard calibration only in the initial steps. Initial

calibration is performed after a calibration reset which completely erases the

calibration and setup memory.

C-Cal – If the last calculated C-Cal values is known, the scale can be calibrated

without weights.

TARE

XQ/G



/R$'

TARE



Blinks
Capacity

1) Remove the seal screw. With

a small screwdriver press the
Cal Switch behind the seal
screw.

The CAL Setup Menu
appears.

2) Press

TARE

to Start the

Calibration Procedure.

3) The display reads UnLd

(unload) indicating you
should remove all weight
from the hook.

4) Ensure the 3460 is

motionless, then press

TARE

. The 3460 sets the

zero calibration point. Wait
while the display is blinking.

5) If the zero is in range, the

scale will display PASS.
Then LoAd1 is displayed.

6) Load the scale with a

precision test weight.

For single

span point calibrations a test weight
of 20% of capacity or more is
recommended.

7) Press

TARE

. The 3460

flashes the capacity. If you
are loading the scale with
the capacity weight, skip to
step 10.

&$/

Standard Calibration Procedure

3$66

No weight
on hook

Flashes
Zero

Lift Test
Weight

example capacity

blinking

TARE

Cal Switch

LO BA

CAL

Remove
Hex Seal
Screw

Press Switch
behind the hole
with a small
Screwdriver

CAL

Figure 5-1. Standard Calibration Procedure