4 calibration setup menu, 5 standard, 4 calibration setup menu 5.5 standard – Rice Lake MSI-3460 Challenger 3 User Manual

Page 38

Advertising
background image

34 MSI-3460 Operator’s Manual

5.4 Calibration Setup Menu

The calibration setup menu contains two items: standard and auto zero maintenance

“Auto0.” In addition, more menus will appear that are transferred from the main setup

menu when Legal for Trade settings are used.

&$/

&&D/

6HWYS

6WDQG

$YWR

LQGYV

+E

U

XQLW

2Q

2II

Cal Standard

Auto Zero

Maintenance

Calibrate with Weights

Calibrate with Constant

Industrial Setting

(default)

Handbook 44, NTEP

R-76, OIML

Industrial, restricted to
1 Weight Unit

AZM On

AZM Off

USER

F

Cal Switch

LO BA

CAL

Remove
Hex Seal
Screw

Press Switch
behind the hole
with a small
Screwdriver

CAL

Figure 5-8. Calibration Setup

5.5 Standard

There are four selections in the standard menu.

1. Industrial (Indu5) - This is the most common setting for the MSI-3460. With the

industrial standard, you have full range zero, access to units switching, filters,

and peak hold.

2. Handbook 44 (Hb-44) - Sets the scale to enable only approved features per the

NTEP HB-44 rules and regulations. Access is denied to peak hold, and the zero

range may be limited. The filter menu is moved to the cal setup menu, so filters

are only accessible through the cal seal.

3. R-76 (r-76) - Sets the scale to enable only approved features per OIML R-76.

Only kg weight units are available. The zero range is limited to 5% (-2+3%

relative to calibrate zero). Net/Gross function is temporary. Once net weight is

established, pushing an F key set for Net/Gross will cause a maximum five

second display of the gross weight. You must clear the tare to display gross

weight constantly. Other metrological aspects are changed to meet R-76

requirements.

4. One Unit (1unit)-The one unit standard is exactly the same as industrial, except

units switching is inhibited. This is useful for metric only countries. Another use

of the one unit standard is to allow the scale to be calibrated in units other than

lb or kg, since conversions are eliminated.

Contact MSI for more information on the standards settings.