0 appendix, 1 troubleshooting, Appendix – Rice Lake MSI-3460 Challenger 3 User Manual

Page 52

Advertising
background image

48 MSI-3460 Operator’s Manual

8.0

Appendix

8.1 Troubleshooting

Problem

Possible Cause

Solution

Display is blank when
POWER button is
depressed.

Discharged battery

Recharge, allow at least four
hours charge.

Defective battery

Replace

Corroded battery or battery
contacts

Clean contacts

Defective switch or circuit
board.

Requires authorized service.

Display does not
function properly or
front panel button does
not function normally
or scale will not turn
off.

Improperly updated software.

Reinstall software

Faulty circuit board

Requires authorized service.

Loose connectors

Requires authorized service.

Scale does not
respond to weight
changes.

Out of calibration

Calibrate

Faulty load cell

Replace

Load cell connector

Check connector and wires.

Display over ranges
below 100% capacity.

Tared weight is added to load
to determine overload point.

Return to gross weight mode.

Zero requires adjustment.

Rezero the scale

Too much weight has been
zeroed.

Rezero the scale

Display Drifts

AZM (Auto0) is turned off.

Turn AZM on

Rapid temperature changes
such as moving the scale
from indoors to outdoors.

Wait until the scale
temperature has stabalized.

Displayed weight
shows large error.

Scale not zeroed before load
is lifted.

Zero the scale with no load
attached.

lb/kg units causing confusion

Select proper units

Requires recalibration

Recalibrate

Display reading not
stable.

Excessive vibration in crane
system.

Increase filtering or increase
‘d’ in cal.

Excessive side loading

Improve load train symmetry

Load cell faulty

Check LC connections

Display toggles
between “Error” and
“Load”

Weight exceeds capacity

Reduce weight immediately

Faulty load cell or wiring

Check LC and LC wiring

Display toggles
between “Error” and
“A2DLo”

A/D is saturated negative

Check LC and LC wiring

Figure 8-1. Troubleshooting