14) The display now reads

LoAd3 or LoAd4.
Repeat steps 12 and 13
for additional span points
press

TARE

or press

ZERO

if you are finished.

After LoAD4 has passed, the 3460
will automatically advance to the next
step.

/RDG

/RDG

or

TARE

To enter another
span point

15) The display shows CAL’d

to indicate that the
calibration was successful.

16) Press

TARE.

The display

flashes C-CAL followed
by the C-CAL number.
Make a note of this number.

If the X1000 annunciator is lit,
also make a note of that.

This number is used to calibrate when
test weights are unavailable. C-CAL
does not replicate multipoint
calibrations.

17) Press

ZERO

to store the

calibration constants, or
after eight seconds, the
calibration constants will
automatically stored. After
the store message is
displayed, the scale goes
to the next item on the cal
menu, SEtuP.

18) Press

ZERO

to exit the Cal

menus and start up the
standard weight display.
The 3460 is ready for use.

19) After ensuring the

calibration is acceptable,
replace the hex seal screw.

If the scale is being used as a Legal-
for-Trade device, place a lead-wire
seal through the cal screw over to
the adjacent screw.

&&D/ 

Example Value only.

VWRUH

ZERO

ZERO

&$/G

6HWYS

TARE

Flashing

ZERO

LO BATT

CAL

X1000

MSI-3460

Model No.

To skip the C-CAL step

28 MSI-3460 Operator’s Manual

Figure 5-3. Standard Calibration Procedure, Continued.