Xq/g, C-cal calibration procedure – Rice Lake MSI-3460 Challenger 3 User Manual
Page 36
TARE
XQ/G
&&$/
TARE
USER
F
1) Remove the seal screw. With
a small screwdriver press the
cal switch behind the seal
screw.
The CAL setup menu
appears.
2) Press the
USER
key to
scroll to the C-Cal menu
selection.
3) Press
TARE
to start the C-
Cal procedure.
If this calibration is using a cleared
3460, the INITIAL CALIBRATION
procedure will start here. You must
set the cal unit, capacity, and count-
before proceding to step 4,
4) The display reads UnLd
(unload) indicating you should
remove all weight from the
hook.
5) Press
TARE
. The 3460 sets
the zero calibration point.
6) If the zero is in range, the scale
will display PASS.
7) Next the display reads
CCAL? to ask that you look
up the C-Cal value to be used.
8) Press
TARE
. The 3460 is
ready for numeric entry of the
C-CAL value.
9) To enter an C-Cal value
press
USER
. The display
far left digit will flash zero
indicating that a number
should be entered.
&$/
C-Cal Calibration Procedure
blinking
3$66
No weight
on hook
Flashes
Zero
Input C-Cal value
blinking
Cal Switch
LO BA
CAL
Remove
Hex Seal
Screw
Press Switch
behind the hole
with a small
Screwdriver
&&D/
USER
F
TARE
blinking
CAL
32 MSI-3460 Operator’s Manual
Figure 5-6. C-Cal Calibration Procedure