Getting Started
User Manual for Secura, Quintix, Practum
Leveling
The balance must be level to ensure precise weighing results. Leveling the balance
compensates for slant or unevenness at the place of installation by twisting the
front leveling feet of the balance.
The balance needs to be re-leveled and adjusted each time its setup location is
changed.
The leveling function is different for the various balance models:
– Electronic level indicator (Secura
®
only)
– Conventional level indicator (Quintix
®
and Practum
®
)
Leveling with Electronic Level Indicator (Secura
®
)
In Secura
®
models, the balance's position is checked with sensors. Instructions on
the display guide the user when leveling the balance.
0.00
Max 2100 g
d=0.01 g
LEVEL
y When
LEVEL
appears in red at the top left of the display, you must level the
balance.
LEVEL
t
To start the leveling function, select the
LEVEL
button.
Leveling
Please turn each foot in the
direction of the arrow as
shown.
y The balance position is shown as an animation of a level indicator on the
display.
t
Twist the front leveling feet of the balance in the specified direction.
Leveling
The balance has been leveled.
y When the level indicator animation turns green, the balance is correctly leveled.
t
To exit the function, select
l
.
t
Complete the leveling by calibrating the balance.