Calibration and Adjustment

User Manual for Secura, Quintix, Practum

Calibration/Adjustment using an Internal Calibration
Weight (Secura

®

and Quintix

®

only)

The balance is fitted with an internal calibration weight in the housing. This
calibration weight is automatically placed on the balance by a motor during
internal calibration and adjustment.

t

Ensure that the balance is stable and the weighing pan is empty.

t

Select

A

to zero the balance.

t

Select the Menu key at the bottom left of the display in the current application.

Components

Statistics

Mixing

Weighing

y The menu appears.

t

Select

the

CAL

button to access the calibration function.

Calibrate balance

CAL-Extern

CAL-Intern

y The

Calibrate balance

window appears.

t

Select

CAL-Intern

.

y Calibration begins.

– The internal weight is automatically placed on the balance in the housing.
– The balance is calibrated and then immediately adjusted.
– The internal weight is automatically removed from the balance.

Date and Time:

14-Jun-2012 13:10

Calibration function:

Internal calibration

Start calibration:

Start: manually

Calibration deviation:

Dev 0.01 g

Adjustment function:

Internal adjustment

Adjustment deviation:

Dev 0.00 g

Report

y A report appears on the display.

The report indicates the deviations identified during calibration.
The report also gives the result of the adjustment process.

t

Close the report window: touch

<

.

y The balance is now calibrated and adjusted.