Density

User Manual for Secura, Quintix, Practum

t

To save the weight value, select

v

.

t

Remove the sample from the weighing pan and place it in the sample holder of

the density determination kit using forceps.

! Ensure that the sample is completely submerged in the buoyancy liquid and

that no air bubbles have formed on the sample.

Max 220 g

d=0.0001 g

4.4787

Weight in medium

END

+

Net

y The weighing display shows the step

Weight in medium

.

t

To save the weight value, select

v

.

Max 220 g

d=0.0001 g

11.86

END

+

y The balance calculates the density of the sample and displays this value.

t

To exit the function, select

x

.

y If the GLP printout function is activated,

the GLP printout is automatically printed (see “Printout" on page 28).