Rice Lake Secura - Quintix and Practum Series - User Manual User Manual
Page 27
27
System Settings (Setup Menu)
User Manual for Secura, Quintix, Practum
Ambient conditions
menu option:
t
Choose whether the conditions at the balance location are
Stable
or
Unstable
.
If
Unstable
is selected, adverse ambient conditions (drafts, vibrations) are
filtered by changing the measurement time for the weight values.
Application
menu option:
t
Choose whether the balance will be used for
Weighing
or
Dosing
.
This setting is used to compensate for load fluctuations on the display.
–
Weighing
: The display responds very quickly to fast load changes.
The display responds more slowly to small changes in weight (in the digit
range). This setting is suitable for normal weighing.
–
Dosing
: The display responds quickly to small changes in weight, making
it faster and more accurate for additional dosing and container filling
functions.
Stability signal
menu option:
t
Choose whether stability is shown with
High accuracy
,
Medium accuracy
,
or
Fast
on the display.
The balance stability is displayed as soon as the weighing result is constant
within a defined range. Until stability is reached, the measured values is shown
in gray on the display and only turns black once the balance is deemed stable.
Zero / Tare
menu option:
t
Define settings for zeroing and taring.
–
Zero/Tare function
: Choose whether the balance is zeroed and tared
With stability
or
Without stability
. If
Without stability
is selected, the
balance will automatically be tared when
D
is selected.
If
With stability
is selected, the balance is tared the next time stability is
reached after
D
is selected.
–
Automatic zero (zero tracking function)
: When this option is active,
changes of a set fraction of scale intervals per second from the display zero
point are tared automatically.
–
Zero/Tare at power on
: When this option is active, the balance is
automatically zeroed or tared when switched on.