Calibration and Adjustment

User Manual for Secura, Quintix, Practum

Automatic Calibration/Adjustment
with isoCAL

Fully automatic calibration/adjustment ensures that the balance is automatically
calibrated and adjusted when predefined time intervals or temperature values are
exceeded.

This function varies in different models:
– Secura

®

: isoCAL is automatically carried out based on time or temperature,

and each time the balance is leveled. This function can be switched off via the
system settings (see “Calibration/Adjustment" on page 26).

– Quintix

®

: isoCAL is automatically carried out based on time or temperature.

This function is always active and cannot be switched off.

h

In

Secura

®

models, every adjustment process, including the identified deviations,

is documented and save on the balance. The saved reports can be viewed via the
system settings and printed out on a PC or Sartorius lab printer (see “Calibration/
Adjustment" on page 26).

0.00

Max 2100 g

d=0.01 g

isoCAL

y If the ambient temperature has changed since the last calibration/adjustment

or the preset time interval has been exceeded, the red

isoCAL

button appears

on the display.

y If the isoCAL function is set to

Automatic

, the calibration/adjustment process

starts automatically.

! No additional error message will be shown when the prompt is ignored for

longer than 60 seconds.

y If the isoCAL function is set to

Info, manual start

, the function must be started

manually.

isoCAL

t

To start the calibration/adjustment function manually, select the

isoCAL

button (Secura

®

, Quintix

®

).

y Calibration begins automatically.

Date and Time:

14-Jun-2012 15:18

Calibration function:

Internal calibration

Start calibration:

Start: isoCAL (Niv)

Calibration deviation:

Dev -0.01 g

Adjustment function:

Internal adjustment

Adjustment deviation:

Dev 0.00 g

Report

y A report appears on the display (Secura

®

and Quintix

®

only).

The report indicates the deviations identified during isoCAL calibration.
The report also gives the result of the adjustment process.

t

Select

<

.

y The balance is now calibrated and adjusted.