ISO/GLP-compliant Printout

User Manual for Secura, Quintix, Practum

ISO/GLP-compliant Printout

Characteristics

The device information, ID, and the current date can be printed before (GLP header)
and after (GLP footer) the values from the weighing series.
The following data is printed out:

GLP Header:
– Date
– Time at the start of a weighing series
– Balance

manufacturer

– Balance

model

– Model serial number
– Balance software version number (BAC)
– Display software version number (APC)
– Identification number (ID)
– Device and batch ID (if activated in the system settings)

GLP Footer:
– Date
– Time at the end of a weighing series
– Field for signature

Configuration

To print the ISO/GLP printout, define the following system settings (see “Printout"
on page 28):

t

Activate ISO/GLP-compliant printouts:

Go to the

Printout

menu, select “Manual with stability,” and then under

Manual print

set the

ISO/GLP printout

option to

On

.

t

Set printout format of date and time:

Go to the

Printout

menu and then under

Manual print

select

Manual print format

. Select any setting except “Value w/o identifier."

h

The ISO/GLP printout is not printed if “Value w/o identifier” is set when
connecting to a PC.

– The ISO/GLP printout is only printed if the “Manual with stability” setting is

selected under Printout.
“Value w/o identifier” is only displayed under “Manual print format” when
printing out to the PC.

Operation

t

To print the header and the first measurement value:

Select the

d

button, or

w

if an application is active.

y The header is included with the first printout.

t

To print the header and reference data automatically when an application

program is active: Select the

x

button.

t

Print the footer and exit the application: Select the

f

button.

Advertising