Getting Started

User Manual for Secura, Quintix, Practum

Leveling with a Conventional Level Indicator

(Quintix

®

and Practum

®

)

L

R

R

L

L

R

. . The.Quintix

®

.and.Practum

®

.models.are.fitted.with.a.conventional.level.indicator ...

This.is.located.to.the.left.of.the.display ..The.position.of.the.air.bubble.in.the.
indicator.shows.whether.or.not.the.balance.is.correctly.leveled .

t

. Twist.both.of.the.front.leveling.feet.of.the.balance.to.move.the.air.bubble.so.

that.it.is.centered.in.the.circular.marking .

The.figure.shows.which.leveling.foot.should.be.twisted.in.which.direction ..
Generally,.both.feet.must.be.adjusted.in.order.to.position.the.air.bubble.exactly.in.
the.center.of.the.circular.marking .

Examples:
–. If.the.air.bubble.is.too.far.to.the.lower.right,.twist.the.right.leveling.foot.

counterclockwise.(see.figure) .

–. If.the.air.bubble.is.too.far.to.the.upper.right,.twist.the.left.leveling.foot.

clockwise .

–. If.the.air.bubble.is.too.far.to.the.left,.twist.the.left.leveling.foot.

counterclockwise.and.the.right.leveling.foot.clockwise .

t

. Check.the.level.indicator.after.each.adjustment.and.correct.if.necessary .

t

. Complete.the.leveling.by.calibrating.the.balance .