Getting Started
User Manual for Secura, Quintix, Practum
Leveling with a Conventional Level Indicator
(Quintix
®
and Practum
®
)
L
R
R
L
L
R
. . The.Quintix
®
.and.Practum
®
.models.are.fitted.with.a.conventional.level.indicator ...
This.is.located.to.the.left.of.the.display ..The.position.of.the.air.bubble.in.the.
indicator.shows.whether.or.not.the.balance.is.correctly.leveled .
t
. Twist.both.of.the.front.leveling.feet.of.the.balance.to.move.the.air.bubble.so.
that.it.is.centered.in.the.circular.marking .
The.figure.shows.which.leveling.foot.should.be.twisted.in.which.direction ..
Generally,.both.feet.must.be.adjusted.in.order.to.position.the.air.bubble.exactly.in.
the.center.of.the.circular.marking .
Examples:
–. If.the.air.bubble.is.too.far.to.the.lower.right,.twist.the.right.leveling.foot.
counterclockwise.(see.figure) .
–. If.the.air.bubble.is.too.far.to.the.upper.right,.twist.the.left.leveling.foot.
clockwise .
–. If.the.air.bubble.is.too.far.to.the.left,.twist.the.left.leveling.foot.
counterclockwise.and.the.right.leveling.foot.clockwise .
t
. Check.the.level.indicator.after.each.adjustment.and.correct.if.necessary .
t
. Complete.the.leveling.by.calibrating.the.balance .